CT2 broken
P201 is no more counting time correctly:
- timer counter returns tick instead of seconds
SESSION_SXM [1]: ct(0.002, p201, save=False)
Tue Feb 11 15:45:30 2020
...
iodet = 652.0 ( 326000.0/s)
sdd2 = 0.0
sec = 25000.0 ( 12500000.0/s) <----------- ☠
NB: commit ed41fc4e9b11ff78f500ca7116a681c0b2e1dbc2
was counting in seconds
- freq specified in YML seems ignored:
yaml:
clock: CLK_100_MHz
server debug:
DEBUG 2020-02-12 21:57:56,050 CardInterface(/dev/ct2_0): ioctl CT2_IOC_DDINT
DEBUG 2020-02-12 21:57:56,050 rpc.CT2: rpc call configure() = None
DEBUG 2020-02-12 21:57:56,054 rpc.CT2: rpc get timer_freq = 12500000.0
DEBUG 2020-02-12 21:57:56,059 rpc.CT2: rpc get internal_timer_counter = 11
DEBUG 2020-02-12 21:58:02,173 rpc.CT2: rpc call __hash__() = -9223363306747575124
DEBUG 2020-02-12 21:58:02,178 rpc.CT2: rpc call __hash__() = -9223363306747575124
Edited by Cyril Guilloud