Update MX sample tracking configuration
- remove
Required resolution"
from diffraction plan
- rename
Aimed resolution
by Initial resolution
- add description (tooltip) and default values
Protein acronym: (Required field) Defines the protein that is registered with the ESRF safety group (SMIS)
Sample name: (Required field) User-defined unique identifier
Pin barcode: Barcode identifier
Experiment type Define MXPressE/O/SAD/Score/M/P/P_SAD/F (Default: MXPressE)
Crystal form: If present (from an associated PDB) used for strategy calculation and autoprocessing (except some Grenades jobs)
Initial resolution: Resolution that the detector will be set to for mesh scans and default data collection or MXPressO (Default: 2.0 Å)
Radiation-sensitivity: BEST input in the case of highly radiation-sensitive or insensitive crystals (0.5–2.0 - low to high sensitivity) (Default: 1)
Required completeness: % of Completeness required (value between1 and100) (Default: 99%)
Required multiplicity: Multiplicity of the final data set (Default: 4)
No. of positions: For multiple crystals on your support or for helical data collection (MXPressP) (Default: 1 (5 for MXPressP and P_SAD)
Beam diameter: Select appropriate beam size for crystals (if blank will be dynamically adapted to crystal size) - value in µm (Default: 30 µm)
Forced Space Group: Select space group for strategy calculation and autoprocessing if cell not known or no PDB (Default: None)
Total rotation angle: Select required rotation range for data collection (Default: MXPressE - minimum required; MXPressE_SAD - 360°; MXPressO - 180°; MXPressF - 180°; MXPressP - 180°/positions; MXPressP_SAD - 360°/positions)